Master-Slave Robot Manipulator for MRI-Guided Needle-Based Abdominal Interventions

arXiv CS · · 1 min read · Engineering & Technology

Read research and analysis on Master-Slave Robot Manipulator for MRI-Guided Needle-Based Abdominal Interventions published by ICANEWS, a global research journal for emerging researchers.

Key Takeaways

  • An MR-safe, master-slave robot manipulator was developed for abdominal interventions in an MRI chamber.
  • The system uses fluid transmission for motion and force between a 2+1-DoF master and slave manipulator.
  • A digital master controller provides multimodal control capabilities.
  • Elastomeric fluid actuators handle remote angulation control.
  • Graphite piston cylinders manage remote needle insertion actuation, offering sub-newton force transparency and sub-millimeter motion.
  • The device supports manual, digital, hybrid, and collaborative control modes.
  • Preliminary in-vivo pig experiment results confirmed functional viability in bedside, manual control mode.

Why This Matters

This technology provides a method for conducting needle-based abdominal interventions with real-time MRI guidance. Its multimodal control capabilities could offer enhanced precision and flexibility for clinicians during complex procedures within the MRI environment.

Overview

A master-slave robot manipulator has been developed for needle-based teleoperation within an MRI chamber, specifically targeting abdominal interventions. The system is designed to be MR-safe and facilitates real-time MRI-guided procedures. It integrates a human-operated master controller with a slave manipulator through fluid transmission, complemented by a digital master controller for multimodal operational capabilities.

Approach

The system architecture comprises a 2+1-DoF (degrees-of-freedom) master controller manipulator that transmits motion and force to a corresponding 2+1-DoF slave manipulator. This transmission is achieved via a fluid-based mechanism. A digital master controller is incorporated to provide control capabilities beyond those typically found in split-axis or mode-switchable hybrid human-digital controller configurations detailed in prior studies.

For remote angulation control, high input impedance, low-leakage, elastomeric fluid actuators are employed. For remote actuation of the needle insertion axis, low-friction graphite piston cylinders are utilized. This choice for the needle insertion axis is based on the system's ability to provide sub-newton force transparency and sub-millimeter motion transmission over bedside fluid piping lengths.

The manipulator supports various control modes, including manual, digital, hybrid, and collaborative. Collaborative tasks demonstrated include assisted tissue penetration, fault-driven virtual fixture, and motion compensation via feedback control.

Findings

Preliminary MR scanner results were obtained from an in-vivo pig experiment. These results demonstrated the functional viability of the manipulator in a bedside, manual control mode configuration.

Why This Matters

The system enables real-time MRI-guided interventions by providing a MR-safe teleoperation platform. This approach integrates manual, digital, hybrid, and collaborative control modes, offering flexibility for complex procedures.

Key Limitations Mentioned by Researchers

The provided information details a preliminary experiment in a specific configuration (bedside, manual control mode) with an in-vivo pig. The extent of testing across all described control modes (digital, hybrid, collaborative) and other potential applications is not elaborated upon in this abstract.